Cataract procedures;Cataract Treatment;Patient database – create cataract treatment ;Capsulotomy:
Enter desired diameter of capsulotomy (normally between 4.8 mm and 5.5 mm)
Height is the height of the capsulotomy cut, usually left as default: 800 μm ? can be adjusted on the fly after the docking
Energy used is 7000 nJ
leave spacings default (6.0 μm/4.0 μm);Lens Fragmentation:
Select fragmentation pattern (radial cuts only, circular cuts only, circular and radial cuts, quadrant and radial cuts)
Energy used is between 7000 nJ to 8000 nJ (higher energy for harder cataracts)
leave spacings default
(10.0 μm/10.0 μm);Radial cuts:
Select number of radials
Enter outer diameter
Circular cuts:
select number of circulars
Select outer diameter (smallest circular always diameter = 500 μm)
Circular and Radial cuts:
Number of radials = 4
Enter outer diameter of radials
Select number of circulars
Enter outer diameter of circulars;Quadrant cuts:
Select number of quadrants
Enter quadrant outer radius
Enter alignment angle α
Enter opening angle β
;Most common used pattern:
Radial cuts (“Pizza cuts”)
On the basis of Divide and Conquer phaco technique
4 radial cuts recommended for very hard cataract/big nucleus:
Because of min. posterior safety distance a hard cataract might not be cut all the way through: be careful with cracking
Best is to complete the cut with the phaco tip
Combination of Radial cuts and circular cuts
Removal of a central cylinder to create some space for further cracking of the lens;GUI – possible lens frag combinations;GUI – cataract ranges and settings;Patient database – select treatment ;GUI Cataract;OCT;After the marking of the pupil the capsulotomy is automatically centered on the pupil
If the pupil is too small to shoot the intended diameter, capsulotomy size will be automatically reduced and a warning message will appear
As there is a safety zone between the marked pupil and the laser area of 0.5 mm the max. capsulotomy size is pupil size decreased by 1 mm
您可能关注的文档
- Herding文献-Herding and anti-herding.pdf
- HDPE 石蜡 活性炭三元相变储能材料的制备及热性能研究 土勋勋.pdf
- Herding文献-Herding and information based trading.pdf
- Herding文献-Herding in humans.pdf
- Herding文献-Herding over the career.pdf
- Guideline文献PRACTICE MANAGEMENT GUIDELINES FOR STRESS ULCER PROPHYLAXIS.pdf
- Herding文献-What drives the hearding behaviour of individual investors.pdf
- Herding文献-Who herds.pdf
- Herding文献-What have we learned from social learning.pdf
- HME文献3 Social identity precursors to the hostile media phenomenon Partisan perceptions of coverage of the Bosnian conflict.pdf
原创力文档


文档评论(0)