GC-MS Screen for the Presence of Melamine and Cyanuric Acid(formerly titled GC-MS Screen for the Presence of Melamine)
(Adapted from FDA/ORA Forensic Chemistry Center SOP T015) Title Revised April 25, 2007
PURPOSE:
This procedure provides a general guide for the sample preparation and analysis of wheat gluten and pet food matrices for melamine using gas chromatography/mass spectrometry.? This procedure is designed to screen the TMS derivative of a methanol extract of the sample and it incorporates instrumental parameters which permit 10.5 minute run times.
SCOPE:
This procedure is applicable to dry wheat gluten, dry pet food, and wet “cuts and gravy” type pet food.? This procedure involves extractions in methanol and therefore preselects compounds which are soluble or partially soluble in methanol.? Due to solubility limitations of melamine in methanol, this method is intended to be a qualitative screen only.
RESPONSIBILITY:
It is the responsibility of the analyst to note any modifications to this procedure in the worksheet.
DEFINITIONS AND ACRONYMS:
TMS – trimethylsilyl
SAFETY CONSIDERATIONS:
Accepted safety measures should be employed when working with chemicals and pressurized gases.
EQUIPMENT:
Agilent 5975 GC-MS system equipped with a 30m DB-5MS column (or equivalent)Reacti-Vap – Evaporating/Heating Module
REAGENTS:
Methanol: HPLC gradePyridine: Certified A.C.S.BSTFA with 1% TMCS: bis(trimethylsilyl)trifluoroacetamide with 1%Trimethylchlorosilane (e.g. Sylon BFT, Supelco)
QC ELEMENTS:
A blank should be run at the onset of each analysis and then randomly throughout the analysis if there is suspicion of carryover.? A fortified sample (see Part C) should be analyzed with each set of samples to demonstrate effective system performance.
PROCEDURES (PROCESS DETAILS):
This procedure may be used with the GC operating in either the splitless or split (1:20) mode.? Generally, the splitless mode is advisable for samples where more sensitivity is required.?? The split
您可能关注的文档
最近下载
- 团知识竞赛试题及答案.docx VIP
- 湖南工程学院《大学物理》2023-2024学年第一学期期末试卷.doc VIP
- 小学社团活动——啦啦操18节课时教案.docx
- 在线网课学习课堂《工程伦理》单元测试考核答案.docx
- 2024年山东省《辅警招聘考试必刷500题》考试题库附完整答案【全优】.pdf VIP
- T_CECS 992-2022CN 预应力压接装配混凝土框架应用技术规程.docx
- 【课件】第1课+美术作品的创作+课件-高中美术人教版(2019)美术鉴赏.pptx VIP
- 网络交易案件办理专题培训课件.pptx
- 《接发列车工作 》课件(下).pptx VIP
- 粤版_通用技术2_技术与设计2__教案.doc VIP
文档评论(0)