Tool bar and button bar (Many SEM functions are mouse-controlled) To insert your sample Check that high voltage (HV) is off (no yellow background) Click on Vent Insert sample. Wear gloves or use tools. Check sample height with “elephant nose”. Bottom of nose is where highest point on sample should be. Click on quadrant showing CCD image (typically lower right) Make sure image is live, not paused! (No double green bars in upper right corner.) Slowly close door while watching monitor to assure that sample/holder is clear of pole piece and detector Click on “Pump” with left hand while holding