* Hysteresis allows the sensed intrinsic rate to decrease to a value below the programmed lower rate before pacing resumes. Hysteresis provides the capability to maintain the patients own heart rhythm as long as possible, while pacing at a faster rate if the intrinsic rhythm falls below the hysteresis rate. The hysteresis rate is always the lower rate limit. The lower rate limit is initiated by a paced event, while the hysteresis rate is initiated by a non-refractory sensed event. In the example above, the lower rate limit is 60 ppm (1000 ms), while the hysteresis rate is 50 ppm (1200 ms).