* Make sure patient is not connected to ventilator. Enter service mode and initiate EST as follows: a.Turn on ventilator. b.While BDU ventilator inoperative indicator is lit, push TEST button within 1 second. DO NOT HOLD BUTTON IN. The ventilator automatically enters the service mode. c.On lower screen, touch EST. If the system does not enter EST, perform an expiratory valve calibration. Install gold standard circuit to ventilator outlet. The ventilator automatically starts the test sequence (Table 1 9). Some tests require your intervention, and will wait indefinitely for your response. Oth