Now also compare the BIER or resistometer cycle to a typical hospital prevacuum cycle that may have anywhere from a 5-10 minute come up time. The point of all this is that sterilization indicators are tested and documented in resistometers (per standards requirements) with precise but different conditions than those in which the indicators are typically used, and, this can have a very great impact on how these indicators perform in actual hospital cycles because there is lethality in the come-up times of these sterilizers. Now also compare the BIER or resistometer cycle to a typical hospital