Artifacts associated with implementation of the Grangeat formula
Seung Wook Leea)
CT/Micro-CT Laboratory, Department of Radiology, University of Iowa, Iowa City, Iowa 52242 and
Department of Nuclear and Quantum Engineering, Korea Advanced Institute of Science and Technology,
Daejeon, 305-701, South Korea
Gyuseong Cho
Department of Nuclear and Quantum Engineering, Korea Advanced Institute of Science and Technology,
Daejeon, 305-701, South Korea
Ge Wangb)
CT/Micro-CT Laboratory, Department of Radiology, University of Iowa, Iowa City, Iowa 52242
Received 29 April 2002; accepted for publication 30 September 2002; published 27 November 2002
To compensate for image artifacts introduced in approximate cone-beam reconstruction, exact
cone-beam reconstruction algorithms are being developed for medical x-ray CT. Although the exact
cone-beam approach is theoretically error-free, it is subject to image artifacts due to the discrete
nature of numerical implementation. We report a study on image artifacts associated with the
Grangeat algorithm as applied to a circular scanning locus. Three types of artifacts are found, which
are thorn, wrinkle, and V-shaped artifacts. The thorn pattern is created by inappropriate extrapola-
tion into the shadow zone in the radon domain. If the shadow zone is filled in with continuous data,
the thorn artifacts along the boundary of the shadow zone can be removed. The wrinkle appearance
arises if interpolated first derivatives of the radon data are not smooth between adjacent detector
planes. In particular, the nearest-neighbor interpolation method should not be used. If the number of
projections is not small, the bilinear interpolation method is effective to suppress the w
您可能关注的文档
- Direct sequencing of the human microbiome readily reveals community differences英文文献.pdf
- Better Naive Bayes classification for high precision spam detection英文文献.pdf
- Automatic Boundary Detection of the Left Ventricle from Cineangiograms英文文献.pdf
- Monitoring multiple distances within a single molecule using switchable FRETRETRET英文文献.pdf
- A short scan reconstruction for cone beam CT using shift invariant FBP and equal weighting英文文献.pdf
- Toward a Simplification of the Protein Folding Problem A Stabilizing Poly alanine a Helix Engineered in T4 Lysozyme英文文献.pdf
- Fluorescence Correlation Spectroscopy. I. Conceptual Basis and Theory英文文献.pdf
- A granulocyte macrophage colony–stimulating factor and interleukin 15 fusokine induces a regulatory B cell population with immune suppressive properties英文文献.pdf
- Corporate social responsibility, partnerships, and institutional change The case of mining companies in South Africa英文文献.pdf
- Pseudoelastic SMA Radius Size Effects on the Damping of Structural英文文献.pdf
文档评论(0)