AK95(95S)英文服务手册-4

Calibrations If the result from the deflation speed test is below the limit, check that the cuff hose is correctly connected to the cuff and that the tube is not kinked or squeezed. Otherwise the failure is caused by the BPM module and it must be replaced and sent to Gambro for repair. If the result from the pressure transducer cali- bration test is outside the limits, it indicates that the self-calibration function within th