蒸汽灭菌和冻干工艺的验证及风险分析.pptx

VALIDATION and RISK ANALYSIS Steam Sterilization and Lyophilizzation ;CONTENT;GMP VALIDATION ;Extracted from the ROSENHEIM Report The AIR was not properly removed from the chamber The drain of the chamber was blocked by pieces of glasses The lower section of the load did not reach the sterilization temperature The temperature recorder showed this anomaly but people thought it was not working properly Sterility tests were carried out only on the upper layers of the load;Validation;Terminal Moist-Heat Sterilization;Validation;CONTENT;What is being Sterilized?;Defining the sterilization process;