1 The device has detected VF, is charging its capacitors for defibrillation, and is confirming VF using a confirmation interval of 460 ms (VT Interval + 60 ms). 2 The device completes charging and starts a 100 ms refractory period while continuing VF confirmation. A refractory event occurs, and the device starts a new confirmation interval. 3 On the second arrhythmic event after charging, the device delivers the defibrillation CV是非约定的放电,需要再确认和同步放电。 充电过程中如有4/5的间期心律失常窗口,则充电放弃;如充电后只要马上出现VS即放弃治疗。 The device confirms the presence of the detected arrhythmia differently after charging than it does