心脏起搏器培训资料:SenseAbility 和附加功能.ppt

* * * * * How low can it go * How low can it go * * * Put all components of SenseAbility together with this slide. * Put all components of SenseAbility together with this slide. SJM internal use ONLY Maximum Sensitivity (The “Floor”) Definition: The most sensitive level to which the device can decay Anything below the maximum sensitivity is not seen by the device Separately programmable (ICD and pacemaker) Nominals: Ventricular 0.3 mV (max is 0.2mV) Atrial 0.2 mV (max is 0.1mV) SJM internal use ONLY 最大感知灵敏度 定义: ICD能衰减到的最敏感水平 任何低于最大感知灵敏度的信号都不能被ICD看见 ICD和起搏器的最大感知灵敏度可分开程控 默认值: 心室0.3mV