- Anytime there are 2 atrial events within a V-V interval, the device may evaluate the possibility that one of these events is a far-field R-wave. - Brady devices typically have a much longer atrial blanking period following ventricular events to eliminate the possibility of sensing far-field R-waves. The Gem DR doesn’t have this long blanking period (remember that atrial events in the PVAB are seen as AR for dual chamber detection), it does have a 30 ms cross-chamber blanking period but no blanking following a VS event. Therefore it is necessary for the device identify these events as far-fi