Model 6308 CT(Chinese) RS485 PROTOCOL
(for firmware version 1.00+)
Version
BAUD : 9600
START : 1
DATA : 8
STOP : 1
PARITY : NONE
Main Protocol
(
ADDRESS
ACKNOWLEDGE
COMMAND
DATA[0]
.
.
.
DATA[n]
ADDRESS = 1 byte, 0-127 must add 128 to indicate as
address
ACKNOWLEDGE = 1 byte, must be equal to 6
COMMAND = 1 byte, valid commands =
(0,3,4,5,6,11,12,13,14,15,16,17,20)
DATA[0]-DATA[n]= depends on PARAMETER
n = number of bytes
II. DATA[0] – DATA[n] FORMAT
IF COMMAND = 0 then DATA[0]..DATA[37] (page 0)
FUNCTION : to get display data during normal operation.
DATA[0] – DATA[5] = CONDUCTIVITY, ASCII FORMAT
Possible readings: ‘+000.0’–’+999.9’,’+00.00’-’99.99’,
‘+0.000’-‘+9.999’,’OVER ‘,’+ LERR‘,’- LERR’,’+ TERR‘,
’- TERR’
“+ TERR” = Temperature OVER
“- TERR” = Temperature UNDER
“+ LERR” = Temp is higher than ccond. Temp limit
“- LERR” = Temp is lower than ccond. Temp limit
DATA[6] – DATA[11] = Temperature Value, ASCII Format
Possible readings: ‘-010.0’ - ’+120.0’,’UNDER ‘,’OVER ‘
DATA[12] – DATA[17] = Analog Out (mA), ASCII Format
Possible readings: ‘+03.00’ - ‘+22.00’, ‘OFF ‘ , ‘FROZEN’ ‘ERROR ’
‘OFF ‘ = Analog system is OFF.
‘FROZEN’ = Analog output is frozen.
‘ERROR ’ = The difference between 4mA 20mA setting is
less than or equal to 0.10 pH.
DATA[18] – DATA[23] = TDS READING ASCII Format
Possible readings: ‘+000.0’–’+999.9’,’+00.00’-’99.99’,
‘+0.000’-‘+9.999’,’OVER ‘,’+ LERR‘,’- LERR’,’+ TERR‘,
‘- TERR’
“+ TERR” = Temperature OVER
“- TERR” = Temperature UNDER
“+ LERR” = Temp is higher than ccond. Temp limit
“- LERR” = Temp is lower than ccond. Temp limit
DATA[24] – DATA[29] = Cell Constant, ASCII Format
Possible readings: ‘0.0000’-‘9.9999’,’10.000’-’99.999’
DATA[30] – DATA[35] = Temperature Coefficient, ASCII
Possible readings:’+00.00’-‘+04.99’
DATA[36] = 1 byte, FLAGS, binary coded
Bit 0 = if 1 then RELAY
您可能关注的文档
- Maya 产品专家级)【DOC精选】.doc
- megui英文参数详解【DOC精选】.doc
- Mechanical Characterization of Dual Phase Austempered Ductile Iron【DOC精选】.doc
- MeGUI参数设置图文对照详解【DOC精选】.doc
- Memory Card 使用【DOC精选】.doc
- mental_ray_for_Maya中文使用手册【DOC精选】.doc
- methodologically【DOC精选】.doc
- Methods to Cut Down Negative Thinking【DOC精选】.doc
- MEP G99-V-CUT改善报告【DOC精选】.doc
- Mesh Modifications (virtools BB模块汉语翻译及用法详解)【DOC精选】.doc
文档评论(0)