Cardiac output measurements using the bioreactance technique in critically ill patients
Fagnoul et al. Critical Care 2012, 16:460
/content/16/6/460
LET TER
Cardiac output measurements using the
bioreactance technique in critically ill patients
David Fagnoul, Jean-Louis Vincent and Daniel De Backer*
Table 1. Patient characteristics
Abstract
Characteristic
n
11
3
Measurement of cardiac output (CO) using minimally
Patients
invasive devices has gained popularity. In 11 patients we
compared CO values obtained using the bioreactance
Cardiogenic shock
technique – a new continuous, totally non-invasive CO
monitor – with those obtained by semi-continuous
thermodilution using a pulm
您可能关注的文档
- Canadian infants' nutrient intakes from complementary foods during the first year of life.doc
- Canada's implementation of the Paragraph 6 Decision is it sustainable public policy.doc
- Canakinumab (ACZ885, a fully human IgG1 anti-IL-1β mAb) induces sustained remission in pediatric patients with cryopyrin-associated periodic syndrome (CAPS).doc
- Cancer and kinases reports from the front line.doc
- Cancer and programmed cell death.doc
- Canadian medical tourism companies that have exited the marketplace Content analysis of websites used to market transnational medical travel.doc
- Canada’s highest court unchains injection drug users; implications for harm reduction as standard of healthcare.doc
- Cancer genomics identifies determinants of tumor biology.doc
- Cancer Carepartners Improving patients' symptom management by engaging informal caregivers.doc
- Cancer and non-cancer health effects from food contaminant exposures for children and adults in California a risk assessment.doc
原创力文档

文档评论(0)