· See most on FSE-IR scan · Small FOV scan, thin slice · Very bony structure, such as foot and knee · Caused by FID of the 180o RF pulses 增大TE可加大FID与180°射频之间的间隔 An anterior spatial SAT excites the tissue beyond the zoom coil and the crusher gradient from the zoom coil becomes ineffective. A 3-coil CTL (CS123) Zoom mode receives a signal from the tissue, excited by the SAT pulse outside the FOV, and wraps into the phase direction. Refer to the graphics below. 水的进动稍快于脂肪中的质子进动。差异:3.5ppm 1.5T:频率=W0=rB0=42.6*1.5=64MHz 差异=3.5*64=220Hz 伪影特点:出现在频率编码方向上。在较低频率的方向出现一条亮带,而较高频率的方向出现一条暗带。往往可见于眼眶、沿椎体终