网站大量收购独家精品文档,联系QQ:2885784924

黑度计校验规程.doc

  1. 1、本文档共3页,可阅读全部内容。
  2. 2、原创力文档(book118)网站文档一经付费(服务费),不意味着购买了该文档的版权,仅供个人/单位学习、研究之用,不得用于商业用途,未经授权,严禁复制、发行、汇编、翻译或者网络传播等,侵权必究。
  3. 3、本站所有内容均由合作方或网友上传,本站不对文档的完整性、权威性及其观点立场正确性做任何保证或承诺!文档内容仅供研究参考,付费前请自行鉴别。如您付费,意味着您自己接受本站规则且自行承担风险,本站不退款、不进行额外附加服务;查看《如何避免下载的几个坑》。如果您已付费下载过本站文档,您可以点击 这里二次下载
  4. 4、如文档侵犯商业秘密、侵犯著作权、侵犯人身权等,请点击“版权申诉”(推荐),也可以打举报电话:400-050-0827(电话支持时间:9:00-18:30)。
查看更多
NJBST Co.,Ltd. QCD-016 Rev.:0 Page:1 of 3 Title: DENSITOMETER CALIBRATION PROCEDURE DENSITOMETER CALIBRATION PROCEDURE 黑 度 计 校 验 工 艺 规 程 CONTROLLED 在控本□ UNCONTROLLED 非在控本□ COPY NO 副本编号: PREPARED BY 编制: DATE 日期: REVIEWED BY 审核: DATE 日期: APPROVED BY 批准: DATE 日期: NJBST Co.,Ltd. QCD-016 Rev.:0 Page:2 of 3 Title: DENSITOMETER CALIBRATION PROCEDURE 1. Scope This procedure covers the calibration of transmission densitometers used to perform radiographic film density measurements and provides a means of evaluating the reliability of transmission densitometers used for the measurement of radiographic film density. This procedure is prepared reference to SE-1079 STANDARD PRACTICE FOR CALIBRATION OFTRANSMISSION DESITOMETERS of ASME Sect.Ⅴ. 2. Apparatus 2.1 Apparatus should consist of the following: 2.1.1 The X-ray film step tablet standard of the National Center of Measuring having at least 5 steps with neutral densities from at least 1.0 through 4.0. The calibration of the step wedge film shall have been verified within the last 12 month by comparision with a National Center of Measuring serialized step tablet. 2.1.2 Transmission Densitometers with digital readout displays specifically manufactured for the purpose of measuring the range of film densities, described in 2.1.1 shall be used. 2.1.3 Manufacturers Operating Instruction for Appropriate Transmission Densitometer. 3. Calibration 3.1 Full-scale linearity calibration of densitometer shall be performed at least every 90 days during use. 3.1.1 Allow a minimum of 30min warm-up time. After this warm-up period, adjust the 0 reading of the densitometer. 3.1.2 Select and position for reading the neutral density closest to 1.0,2.0,3.0,4.0 on the calibrated step tablet. Read and record the density for each step. 3.1.3 Compare the densities recorded with the actual density value

文档评论(0)

zzqky + 关注
实名认证
内容提供者

该用户很懒,什么也没介绍

1亿VIP精品文档

相关文档