高分辨率谱仪的最小噪声共享基线恢复
Nuclear Instruments and Methods in Physics Research A 482 (2002) 441–448
Shared baseline restoration at minimum noise for high
resolution spectroscopy
A. Geraci, I. Rech, E. Gatti*, G. Ripamonti
Politecnico di Milano, Dipartimento di Elettronica e Informazione, Piazza Leonardo da Vinci 32, Milano 20133, Italy
Received 1 July 2000; accepted 12 July 2001
Abstract
Optimum filter theory in Nuclear Spectroscopy is revisited, taking into account the noise contribution o fthe baseline
estimator (baseline restorer or other suitable filter). An overall optimization o fthe signal-to-noise ratio is performed,
and new filters are obtained. They include as particular cases the cusp filter and the Deighton filter. The theory is carried
out by considering only series voltage noise, but it can easily be expanded to include parallel current noise. r 2002
Elsevier Science B.V. All rights reserved.
Keywords: Baseline restoration; High resolution spectroscopy
1. Introduction immune to baseline level shifts [3]. The second
consists in using temporally distinct filters: one for
In any spectroscopy system, signals at the output measuring at best the pulse amplitude, the other(s)
o fthe shaper amplifier are always overlapped to a for measuring at best the baseline level in time
voltage baseline level. This level is not stable and interval(s) where no pulse are present. Then the
needs to be estimated and subtracted from the filter outputs are weighed and summed up so that
signal, in order not to reduce accuracy o fthe the resulting overall filter b