PB7200呼吸器Service Manual Ch13
SECTION
SECTION
Repair: Exhalation Compartment 13 13
The exhalation compartment may be either a bottom-loading or a front-loading
configuration. Either configuration may have an internal exhalation valve within this
compartment or an external valve as part of the patient tubing circuit.
This section includes removal and installation procedures for the exhalation compartment
assembly. Before repairing the exhalation compartment assembly, familiarize yourself with
the warnings, cautions, and general repair instructions in Section 8.
Warning
To prevent personal injury and equipment or property damage, ensure that electrical power
sources and pneumatic sources are disconnected from the ventilator before servicing.
13.1 Exhalation Bacteria Filter
The exhalation filter helps prevent bacteria that may be present in the exhaled gas from being
vented to room air. It also reduces cross-contamination of the ventilator. A Puritan Bennett
filter captures particles as small as 0.3 µm at a flow of 100 lpm with 99.97% efficiency.
Caution
Always follow recommended procedures for maintaining the exhalation filter (Section 4)
and removing moisture from the patient circuit. Debris or moisture contacting the Q3/T3
sensing element may destroy Q3/T3.
13.1.1 In Bottom-Loading Exhalation Compartment
Remove the exhalation bacteria filter from the