* Note: AT Intervention therapies are programmed from the Arrhythmia Intervention window available via the Parameters screen. All are nominally Off. Atrial Preference Pacing (APP) promotes a consistent activation sequence to address relative bradycardia and suppress the ectopic beats that can trigger atrial tachyarrhythmias. When APP is programmed On, the device adjusts the pacing escape interval to provide continuous atrial pacing slightly higher than the patient’s intrinsic rate. The device: Decreases the pacing escape interval (increases rate) by the programmed decrement (30 to 150