Particle Size Distribution Analyzer AN154
Applications Note Dry Method Development
Applications Note
Applications Note
DRY METHOD DEVELOPMENT FOR
LASER DIFFRACTION PARTICLE SIZE MEASUREMENTS
Developing an appropriate method for measuring particle size distribution
of powders dispersed in air using laser diffraction requires a structured
approach. The basic goals for developing a dry method include: place a
representative sample into the analyzer, disperse the sample using a
pressure that breaks up agglomerates but not individual particles, choose
appropriate system settings for measurement, and test for reproducibility.
Introduction • Accepted standards stress the
Developing an appropriate method for importance of sampling. ISO 13320
measuring particle size distribution (3) and USP 429 (4) both
using laser diffraction requires a advise that a representative
structured approach. Samples sample be prepared using a sample
analyzed as a dry powder have a few splitting technique. (5)
unique considerations compared to • One of the goals of proper method
wet analysis, but the majority of the development is to minimize the
content in this document will be total error. If sampling is ignored
applicable to all methods. the developer doesn’t know which
portion of the total error comes
In all cases the basic goals remain the from the sampling.
same; decide what the purpose of the
measurement is, place a Even once a representative sample has
representative sample into the been placed