Designation:E1106–86(Reapproved2002)e1

StandardMethodfor

PrimaryCalibrationofAcousticEmissionSensors1

Thisstandardisissuedunderthe?xeddesignationE1106;thenumberimmediatelyfollowingthedesignationindicatestheyearof

originaladoptionor,inthecaseofrevision,theyearoflastrevision.Anumberinparenthesesindicatestheyearoflastreapproval.A

superscriptepsilon(e)indicatesaneditorialchangesincethelastrevisionorreapproval.

e1NOTE—Editoriallyreplacedtheterm“inspection”with“examination”in6.2inJune2002.

1.Scope

understood.Ifthiscanbeestablished,thestatedaccuracy

shouldapplyoverthefullfrequencyrangeupto1MHz.

1.1Thismethodcoverstherequirementsfortheabsolute

calibrationofacousticemission(AE)sensors.Thecalibration

yieldsthefrequencyresponseofatransducertowaves,ata

surface,ofthetypenormallyencounteredinacousticemission

work.Thetransducervoltageresponseisdeterminedatdis-

cretefrequencyintervalsofapproximately10kHzupto1

MHz.Theinputisagivenwell-establisheddynamicdisplace-

mentnormaltothemountingsurface.Theunitsofthe

calibrationareoutputvoltageperunitmechanicalinput(dis-

placement,velocity,oracceleration).

1.2Thisstandarddoesnotpurporttoaddressallofthe

safetyconcerns,ifany,associatedwithitsuse.Itisthe

responsibilityoftheuserofthisstandardtoestablishappro-

priatesafetyandhealthpracticesanddeterminetheapplica-

bilityofregulatorylimitationspriortouse.

N

OTE

1—Thestatedaccuracyappliesonlyifthetransferstandard

returnstoquiescence,followingthetransientinput,beforeanywave

re?ectedfromtheboundaryofthecalibrationblockreturnstothetransfer

standard(;100μs).Forlowfrequencieswithperiodsontheorderofthe

timewindow,thisconditionisproblematicaltoprove.

4.2ApplicationsSensors—Thismethodmayalsobeused

forthecalibrationofAEsensorsforuseinnondestructive

evaluation.Someofthesesensorsarelesswellbehavedthan

devicessuitableforatransferstandard.Thestatedaccuracyfor

suchdevices