BC-5180全自动血液细胞分析仪操作规程(国外英文资料)
BC-5180全自动血液细胞分析仪操作规程(国外英文资料)
The bc-5180 fully automatic blood cell analyzer operation procedure
The name and model of the instrument
Instrument name: fully automatic blood cell analyzer
Version 1.2 of BC - 5180
2 manufacturers
Shenzhen minui bio-medical electronics co., LTD
3 test range
For the clinical test, the blood cell count white blood cell counts the hemoglobin concentration
4 detection principle
4.1 overview
This analyzer USES the coulter principle to detect the number of white blood cells, basophilic cells, red blood cells and platelets and volume distribution. Using a colorimetric method to measure the concentration of hemoglobin; Four classifications of white blood cells were obtained by using semiconductor laser flow cell technology. Based on this, the other parameters are calculated by the analyzer.
4.2 sample
According to the configuration, this analyzer provides two kinds of methods: closed in sample mode and automatic method. In this case, the enclosed sample supports the whole blood mode and the pre-dilution mode, and the auto-injection mode supports the whole-blood pattern.
In a closed sample of whole blood work mode, the analyzer will suck the peripheral blood samples and 20 ul 180 ul diluent in closed mixed, forms the dilution ratio of 1:10 dilution sample, then this will be diluted sample to the analyzer samples. At this point, the analyzer will inhale the diluted sample of 80uL (CBC + DIFF).
The analyzer will absorb the 20uL (CBC + DIFF) or 15uL (CBC mode) blood samples in an automated approach.
4.3 dilution sample
Sample under test is divided into two after the drain to the host, in the parallel dilution process through different reagents, respectively form is used for red blood cells, platelets, white blood cell count, hemoglobin and WBC classification measurement of test sample.
According to different needs, the analyzers provide two modes of work - the all-blood working mode and the pre-diluted working mode
4.3.1 full blood mode
White bl