水泥 文本文档(国外英文资料) FBT-9 digital instrument for measuring specific surface area of gas permeability and test procedure of cement specific surface area instrument: 4.4.1. on the power supply, the monitor shows the ambient temperature at that time. After the boot, it should be preheated for 15 minutes. When the temperature sensor is stable, the test can be carried out. 4.4.2 leak check: The barometer of distilled water was added to the bottom line (water line), with a rubber plug will barometer right end mouth plugged, press the keyboard on the leak check button, the instrument automatically open the pump and solenoid valve, when the liquid level gauge body to the top first moment line, the first line at the photoelectric switch, three seconds after the pump and the valve is closed, the liquid level to stop rising. If the system is not leaking, the liquid level remains unchanged. After a minute, the monitor shows good. On the contrary, if the system leaks, the liquid level drops, and when the liquid level drops, passing through the photoelectric switch at the first line, the monitor displays a bunch of cursor, that is, the system leaks. In case of this, it can be solved by checking the connections at each pipe and sealing them with piston grease. No matter whether it leaks or not, after checking, press the confirm button to make the instrument return to the display temperature condition. 4.4. 3: calibration (if the test sample for cement, do not calibrate as long as the input button, and do not enter a password, followed by the confirmation of various parameters, and then will be key to check the corresponding test sample density and void ratio of input instrument directly after 4.4.4) The calibration procedures are as follows: (in the case of an example of calibration) users are not allowed to enter the calibration status, otherwise they will be replaced by the calibrated data at the factory. Note 3: calibration process, in addition to the ambient temperature and liqui

文档评论(0)

f8r9t5c + 关注
实名认证
内容提供者

该用户很懒,什么也没介绍

版权声明书
用户编号:8000054077000003

1亿VIP精品文档

相关文档