波科介入 FilterWire - 中文(保护伞).ppt
* * Achieve Good Guide Catheter Support: Loss of guide catheter engagement may cause accidental filter movement. Ensure Sufficient Landing Zone: Allowing extra room for small inadvertent filter movement is good practice to prevent the filter from coming back into the stent or lesion. Also, be careful not to jam the tip of the balloon or SDS catheter onto the filter’s catheter stop, as this may cause wedging and wire lockup, leading to unintentional filter retraction. Minimize Wire Movement: Secure the filter loop position throughout the procedure to prevent inadvertent movement. In particu