能量色散型X射线荧光分析装置EDX-GP安装条件确认书-岛津

ZEQV-2031 能量色散型X 射线荧光分析装置 EDX-GP 安装条件确认书 岛津企业管理(中国)有限公司 大型分析仪器事业部 目 录 1. 前 言 ··············································································································· 3 2. 实验室条件 ·········································································································· 3 3. 设置例 ················································································································ 4 4. 设 备 ·················································································································· 6 4.1 所需电源········································································································  4.2 检测器的冷却剂 (液氮)···················································································  5. 其 他 ·················································································································· 8 6. 安装前确认表········································································································ 9 - 2 - 1. 前言 感谢您购买岛津能量色散型 X 射线荧光分析装置 EDX-GP。 EDX-GP 我公司在多年研制、生产波长色散型 X 荧光分析仪器基础上,开发的新仪器。本资料是为 了协助您顺利的安装EDX-GP ,在安装工作进行前,需要提前进行准备的工作内容。 为确保仪器稳定,可以进行高可靠性的分析,也为了确保能够长期使用,请您仔细阅读本设置要领 书,并按照其要求进行准备。 2.实验室条件 (1)温 度 10℃~30℃ ・室温变动范围:在 10℃~30℃范围内,变化±5℃以内 ・温度的变化小于 2℃/小时,避免温度的急速变化。 (2)湿 度 40 %~70%(不结露) (3) 空 调 管 理 24 小时 ・空调出风口不要直接吹向仪器 若不能满足以上(1),(2),(3)所述的条件,会发生以下的故障。 1.不进行室内温度的管理 温度急速发生的话(室内温度若