Shutting down the system
Buffers left in the system can precipitate and damage instrument components.
Buffers dissolved in water may precipitate
You may want to shut down the system between analyses, overnight or for a weekend, or for
72 hours or more.
Tip: If you are using Empower software to control the system, set system shutdown
parameters in the Instrument Method Editor. Consult the Empower Help for more information.
If you are using MassLynx software to control your system, set system shutdown parameters
in the Shutdown Editor. Consult the MassLynx Help for more information.
Between analyses
To shut down the system between analyses:
1. Between analyses, continue to pump the initial mobile phase mixture through the column.
This maintains the column equilibrium necessary for good retention time reproducibility.
2. If a few hours will pass before the next injection, slow the flow rate in the interim to a few
tenths of a mL/min to conserve solvent.
Tip: Ensure that Auto-Shutdown for your shutdown method is deactivated.
3. Keep the detector operating, and the column manager at operating temperature, during
this period.
Overnight or weekends
To shut down the system overnight or over a weekend:
1. Flush the column with 1:1 mixture of methanol/acetonitrile. This maintains the column bed
in an active, wetted state.
Requirement: If you are using buffers, you must first flush the column with a high water
content mobile phase (90% H2O). Then stop the pump flow.
2. If possible, extinguish the detector lamp to lengthen lamp life.
3. The column manager can operate overnight but should be shut down over the weekend.
Caution: The power switch on each system instrument controls the basic
operational state of that instrument. A portion of the instrument remains powered-
on after the instrument is switched off. To completely interrupt power to a system
instrument, set the power swit
您可能关注的文档
- 3355 TQS 超高效液相色谱串联质谱 2014-5-9.docx
- NuGenesis LMS安装和配置指南.pdf
- QDa 检测器概述和维护指南.pdf
- 沃特世仪器说明书沃特世仪器说明书masslynx安装指南.doc
- 沃特世仪器说明书QDa QC应用文集.pdf
- 沃特世仪器说明书Replace 2424 nebulizer.pdf
- 沃特世仪器说明书Replace 2424 Lamp.pdf
- 沃特世仪器说明书Rebuild 2998 flow cell.pdf
- 沃特世仪器说明书Plunger Seal Wash.pdf
- 沃特世仪器说明书Performance Maintenance.pdf
文档评论(0)