人教版九年级物理专题
A Numerical Study of the Field Dependence
of Signal-to-Noise Ratio in High-Field MRI
M. E. Kowalski and J.-M. Jin
Center for Computational Electromagnetics
Department of Electrical and Computer Engineering
University of Illinois at Urbana-Champaign
Urbana, Illinois 61801-2991
I. Introduction
The increasing interest in the use of magnetic resonance imaging (MRI)
for quantitative spectroscopy, real-time imaging, magnetic resonance angiog-
raphy (MRA), and other emerging applications has led to increased effort
towards maximizing the available signal-to-noise ratio (SNR) from magnetic
resonance (MR) systems. A necessary development to enabling a designer
to maximize the available SNR is the emergence of an accurate method for
its analysis. However, many of the trends in hardware design intended to
increase SNR (such as the move towards higher Bo-field strengths) make the
analysis of SNR more complicated than in traditional clinical MR systems.
Edelstein e t al. [l]pointed out that there is a quantity, termed the intrin-
sic SNR (ISNR), which represents a fundamental limit of the performance
of an MR system for a given patient/coil configuration. This work, howev-
er, used quasi-statics to describe the -field distribution of the transmitter
coil and the sensitivity field of the receiver coil. While this is an accurate
approach at the frequencies then being considered (up to 21 MH