PB10酸度计标准操作规程中英文对照
1.0 Purpose: 目的
1.1 建立 PB-10 酸度计的使用、维护与保养规程,确保测定数据的准确性。
For establish the use and maintenace on pH Meter PB-10 procedure, to ensure the accuracy of
measurement data.
2.0 Scope:范围
2.1 适用于原料、辅料、中间产品、成品、水质以及缓冲溶液的pH值测定。
Apply to determine the pH value of raw materials, auxiliary materials, intermediate products
and finished products, water quality and buffer solution.
3.0 Responsibility:职责
3.1 实验操作员
Laboratory technician
4.0 Definitions: 定义
4.1 pH值:pH值是衡量一种溶液酸度或碱度的尺度,用下面的公式表示:
pH=-log[H+]
[H+]表示溶液中氢离子的浓度,pH值有时也被称为“氢离子指数”。
pH: pH is a measure of the acidity or alkalinity of a solution and can be represented by this
equation:
pH=-log[H+]
with [H+] representing the concentration of hydrogen ions in the solution. pH is sometimes
referred to as the power of the hydrogen ion in a solution.
5.0 Procedure:规程
5.1 准备
Preparation
5.1.1将变压器插头与 pH计 Power(电源)接口相连,并接好交流电。
Connection power cable to meter power jack and to AC power source.
5.1.2 将 pH复合玻璃电极与 BNC(电极)和 ATC(温度探头)输入孔连接。
Connect a glass pH electrode to the input and ATC connectors.
5.1.3 按 Mode键设置 pH模式(pH,mV或相对 mV)。
Press Mode until the display indicates the appropriate measurement mode (pH or mV/relative
mV ).
5.2 校正
Standardization
5.2.1按 SETUP键,显示屏显示 Clear,按 Enter键确认,清除之前的校准数据,
并选择您所需要的缓冲液组。
Press Setup and the meter displays a flashing Clear icon. Press Enter to clear all existing
buffers and to select a new set of buffers.
5.2.2 将复合电极用蒸馏水清洗,滤纸吸干后浸入第一种缓冲液(6.86),充分
搅拌,按 Standardize键, pH计识别出缓冲液并将闪烁显示缓冲液值。达到稳
定状态后,或通过按 Enter键,测量值即已被保存;pH计显示电极斜率为 100%。
Rinse electrodes with distilled water and blot-dry, place electrodes in the first buffer
solution (6.86), stir. Press Standardize, the meter recognizes the buffer and flashes a buffer icon.
When the signal is stable, or when you press Enter, the buffer is entered. The meter displays
the %slope of the electrode as 100%.
5.2.4将复合电极