罗氏试剂盒说明书英文原版 Troponin T hs STAT 05092728 190 V8.0.pdf

ms_05092728190V8.0 Troponin T hs STAT Troponin T hs (high sensitive) STAT (Short Turn Around Time) REF SYSTEM Elecsys 2010 cobas e 411 190 100 cobas e 601 cobas e 602 English The Elecsys Troponin T hs STAT assay employs two monoclonal Intended use antibodies specifically directed against human cardiac troponin T.40,41 The Immunoassay for the in vitro quantitative determination of cardiac antibodies recognize two epitopes (amino acid position 125‑131 and troponin T in human serum and plasma. This assay can be used as an aid 136‑147) located in the central part of the cardiac troponin T protein, which in the differential diagnosis of acute coronary syndrome to identify necrosis, consists of 288 amino acids. e.g. acute myocardial infarction. The test is further indicated for the risk The Troponin T hs STAT calibrators (Troponin T hs STAT CalSet) contain stratification of patients presenting with acute coronary syndrome and for recombinant human cardiac troponin T (rec. hcTnT). The rec. hcTnT is cardiac risk in patients with chronic renal failure. The