起搏器程控小讲座PPT

Dual-chamber pacing requires attention to these parameters: Lower rate AV and V-A intervals Upper rates Refractory periods Blanking periods In order to provide optimal hemodynamic benefit to the patient, dual-chamber pacemakers strive to mimic the normal heart rhythm. In dual-chamber pacemakers, the lower rate is the rate at which the pacemaker will pace the atrium in the absence of intrinsic atrial activity. Similar to single-chamber timing, the lower rate can be converted to a lower rate interval (A-A interval), or the longest period of time allowed between atrial events. The SAV is usually programmed to a shorter duration than the PAV to allow for the difference in interatrial conduction time between intrinsic and paced atrial events. Think of the difference in the activation sequence between a cycle initiated with an intrinsic atrial event versus a paced atrial event. The cycle starting with the intrinsic atrial event will use the normal conduction pathways between the right atrium and the left atrium. The cycle starting with the paced atrial beat will not use the normal interatrial conduction pathways but will instead use muscle tissue, which takes a little longer to reach the left atrium and causing it to contract. If the AV interval is timed to allow the appropriate amount of time for left ventricular filling when the cycle is initiated with a sensed atrial event, the same duration for the PAV may not be the appropriate amount of time to allow for left ventricular filling when the cycle is initiated by a paced atrial event. Proper LA-LV timing promotes left ventricular filling (atrial kick) and prevents regurgitant flow through an open mitral valve. Therefore, it is beneficial to have separately programmable PAV and SAV intervals. In this example, the lower rate interval is terminated by a sensed atrial event, which initiates a SAV interval (and restarts the the lower rate interval). The A-V interval is employed to allow the appropriate amount of time

文档评论(0)

djdjix + 关注
实名认证
内容提供者

该用户很懒,什么也没介绍

1亿VIP精品文档

相关文档