贴片式动态心电记录仪在临床教学中可行性研究-newcentury
/index.php/jenc Journal of Education in New Century,2016,Vol.4,No.1
Studies on the feasibility of patch type dynamic ecg
recorder in clinical teaching
1 2 1 1 3
Gui-hua Jian , Jin-gen Jiang , Meng Wei , Nian-song Wang , Ju-kang Gu
1The sixth peoples hospital affiliated to Shanghai jiaotong university, Shanghai, China
2Zhongshan hospital affiliated to fudan university, Shanghai, China
3The first peoples hospital of Shanghai, Shanghai, China
Received: May 30, 2016 Accepted: Jun 12, 2016 Published: Jun 31, 2016
DOI:10.14725/jenc.v4n1a1515 URL:/10.14725/jenc.v4n1a1515
This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use,
distribution, and reproduction in any medium, provided the original work is properly cited.
Abstract
Objective: To explore effectiveness of clinical application of the patch type dynamic ecg recorder by comparing with conventional
dynamic ecg recorder Holter. And the clinical ecg teaching models is discussed in this paper. Methods: Using parallel controlled trial
method, 60 patients in two hospitals received Holter examination, and patch heart testing at the same time. A comparative analysis of
two kinds of inspection methods on testing the average heart rate, supraventricular premature beats and ventricular premature beat was
made. Results: The average heart rate of patch type ECG and Holter were 71.87 ±9.28 and 72.55 ±9.06 (P = 0.0879). Two kinds of
detection methods, chamber on the coincidence rate of premature beat (PPS) was 98.33%, 95% CI (91.06, 99.96), Kappa = 0.9462, 95%
CI (0.84, 1.00), good consistency. The accuracy of ventricular premature beat (PPS) was 98.33%, 95% CI (91.06, 99.63), Kappa =
0.9850, 95% CI (0.90, 1.00), good consistency. Patch type heart all platform is suitable for self-study and clinical care study, and
pro