REV. A 1 OF 1
Flushing the 2475 Flow Cell
When to Perform
• Flush and passivate the flow cell after each shutdown, or when dirty. A dirty flow cell
can cause baseline noise, decreased sample energy levels, calibration failure, and
other problems with the operation of the 2475 Detector.
Parts Required
• None
Tools Required
• A wrench suitable for removing and replacing the column
• A solvent miscible in both the mobile phase and water (methanol is commonly used)
• HPLC-grade water
• To passivate the flow cell, a strong cleaning solvent suitable for your system (6N nitric
acid is commonly used)
• A separate container for acid waste
• A means for measuring the pH of acid effluent (if you use an acid as your cleaning
solvent)
NOTE: There is no “Flushing the 2475 Flow Cell” video.
Flush and Passivate the Flow Cell
1. Stop mobile phase flow, then remove the column.
2. Replace the column with a union or section of tubing.
3. Attach tubing from the detector to the injector outlet where the column was
removed.
4. Flush the mobile phase from the detector with a miscible solvent and water.
5. Flush the detector with HPLC-quality water to remove contaminants from the flow
path.
6. Pump 6N nitric acid through the flow cell to clean the internal pathways.
7. Flush again with HPLC-quality water until the pH is neutral.
8. Reattach the column to resume flow of the mobile phase. Use an intermediary
solvent if you are using a mobile phase that is not miscible in water.
NOTE: Be certain that you do not exceed the pressure limit of the flow cell. Set the pres-
sure limit switch on the e2695 Separations Module accordingly for the specific flow
cell being used.
Flushing the 2475 Fl
您可能关注的文档
- 3355 TQS 超高效液相色谱串联质谱 2014-5-9.docx
- NuGenesis LMS安装和配置指南.pdf
- QDa 检测器概述和维护指南.pdf
- 沃特世仪器说明书沃特世仪器说明书masslynx安装指南.doc
- 沃特世仪器说明书QDa QC应用文集.pdf
- 沃特世仪器说明书Replace 2424 nebulizer.pdf
- 沃特世仪器说明书Replace 2424 Lamp.pdf
- 沃特世仪器说明书Rebuild 2998 flow cell.pdf
- 沃特世仪器说明书Plunger Seal Wash.pdf
- 沃特世仪器说明书Performance Maintenance.pdf
- 沃特世仪器说明书Mobile Phase Considerations.pdf
- 沃特世仪器说明书e2695 Syringe Replace.pdf
- 沃特世仪器说明书e2695 Seal Pack Rebuild.pdf
- 沃特世仪器说明书e2695 Sample Loop Replace.pdf
- 沃特世仪器说明书Rebuild 2489 flow cell.pdf
- 沃特世仪器说明书e2695 Lower Needle-wash Frit Replace.pdf
- 沃特世仪器说明书Replace 2414 relief vlv.pdf
- 沃特世仪器说明书e2695 Replace all Seals.pdf
- 沃特世仪器说明书e2695 Inline Filter Replace.pdf
- 沃特世仪器说明书e2695 Check Valve Replace.pdf
文档评论(0)