新生儿气囊活瓣面罩复苏器的监测设备EvaluationoftheAugmented
增强婴儿复苏器的评价:新生儿气囊活瓣面罩复苏器的监测设备
Evaluation of the Augmented Infant Resuscitator: A Monitoring Device for
Neonatal Bag-Valve-Mask Resuscitation
* *† *
Bennett, Desmond J. BA ; Itagaki, Taiga MD ; Chenelle, Christopher T. BA ;
† *†
Bittner, Edward A. MD, PhD ; Kacmarek, Robert M. PhD, RRT
Anesthesia Analgesia: 2018 126 947 –955
背景:每年有 600 万新生儿需要气囊活瓣面罩复苏,提供即时反馈有可能提高复
苏质量。增强婴儿复苏器(AIR)是一种实时反馈设备,旨在识别气囊活瓣面罩
复苏期间的泄漏、阻塞,以及不适当的呼吸频率。但是,它的功能尚未进行评估。
方法:将其连接在呼吸机和呼吸机测试仪之间来测量 AIR 的阻力。为了测试设备
在训练和临床应用的可靠性,将其置于气囊或呼吸机与新生儿人体模型和临床肺
模型模拟器之间。肺模型模拟器模拟 3 种尺寸(2,4 和 6 kg)的新生儿。评估泄
漏,阻塞和呼吸频率的变化。
结果:流量为 5L/min 时,AIR 上的压力降低仅 0.38 cm H2O,该设备对呼吸机呼
吸参数几乎没有影响。在人体模型试验期间,它能够检测到所有泄漏和阻塞,100%
及时正确显示警报。在模拟的临床试验中,AIR 在 6 公斤新生儿模型中效果最佳,
其次是 4 公斤模型,最后是 2 公斤模型。在所有 3 种临床模型中,模型显示 73.5%
的正确指标,当这样做时,耗时 1.6±0.9 秒。
结论:AIR 是可能改善新生儿复苏的一项有希望的创新,但它仅引入阻力边缘值,
并仅在新生儿模型上表现良好,在临床使用前仍需改善其固件。
(马益梅译潘艳、薛张纲校)
BACKGROUND Annually, 6 million newborns require bag-valve-mask
resuscitation, and providing live feedback has the potential to improve
the quality of resuscitation. The Augmented Infant Resuscitator (AIR),
a real-time feedback device, has been designed to identify leaks,
obstructions, and inappropriate breath rates during bag-valve-mask
resuscitation. However, its function has not been evaluated.
METHODS The resistance of the AIR was measured by attaching it between
a ventilator and a ventilator tester. To test the devices reliability
in training and clinical-use settings, it was placed in-line between a
ventilation bag or ventilator and a neonatal manikin and a clinical lung
model simulator. The lung model simulator simulated neonates of 3 sizes
(2, 4, and 6 kg). Leaks, obstructions, and respiratory rate alterations
were introduced.
RESULTS At a flow of 5 L/min, the pressure drop across the AIR was only
0.38 cm H2O, and the device had almost no eff
您可能关注的文档
最近下载
- 小学数学人教版(2025)五年级下分数的意义课件(共21张PPT).pptx VIP
- 《铁路货物站台面及堆货场铺面》壹站(02)8029.pdf VIP
- 小学数学人教一年级上册以内的进位加法旦旦数鸡蛋(区一等奖).pptx VIP
- 安全生产法律法规完整版全套PPT教学课件.pptx VIP
- 教你三步查看老婆微信聊天记录.docx VIP
- 语文九年级下册第三单元课外古诗词诵读《浣溪沙身向云山那畔行》PPT课件 [修复的].pptx VIP
- 2025广西专业技术人员公需科目培训考试答案(87分);广西“一区两地一园一通道”建设;人工智能时代(1).pdf VIP
- 部编小学语文六年级上册第二单元《革命岁月》集体备课教材分析解读.pptx VIP
- 2025广西公需科目培训考试答案(90分)一区两地一园一通道建设人工智能时代的机遇与挑战.docx VIP
- 2024年海南省海口市中考数学试题卷(含答案解析).docx
文档评论(0)